To avoid losing IMAGE-COM measurements, try to complete all measurements on the modality before the reading workflow begins. Using the Reload modality SR button resets all measurements.

Issue

Measurements added to the IMAGE-COM worksheet during reporting are lost when the Reload modality SR button is used.

Details

This issue can occur if a technician continues to work with a study on the modality (adding additional measurements) after the study has been sent to Enterprise Imaging. If the reading is started and measurements are added to the IMAGE-COM worksheet (and not yet exported to the report panel), those measurements are lost when the additional measurements sent from the modality are refreshed in IMAGE-COM using the Reload modality SR button.

Solution

If possible, ensure all the measurements on the modality have been completed before starting a reading using Enterprise Imaging and making measurements in IMAGE-COM. If measurements are lost after clicking the Reload modality SR button, the measurements must be made again.