Window level presets for CT data: ReferenceTopic number: 1501609936566
Two types of window level presets are available for CT data: linear and filtered. A linear preset is defined by center and width values and a filtered preset is defined by a combination of an available linear preset and an image filter.
Window level presets are defined by the following criteria:
Preset | Description |
---|---|
Name | The name of the defined preset. |
Window Center | A numeric value defining the center of the window. This value can be positive or negative. |
Window Width | A numeric value defining window width. This value must be positive. |
The following presets are defined by default for CT data:
Name | Window Width | Window Center | Type |
---|---|---|---|
Lung | 1500 | -500 | Linear |
Mediastinum | 150 | 88 | Linear |
Soft tissue | 350 | 50 | Linear |
Liver | 150 | 85 | Linear |
Bone | 2000 | 500 | Linear |
Abdomen/Pelvis | 300 | 70 | Linear |
Brain | 80 | 20 | Linear |
Post Fossa | 125 | 47 | Linear |
Bone + sharpen filter | 2000 | 500 | Filtered |
Lung + sharpen filter | 1500 | -500 | Filtered |
Soft tissue + smooth filter | 350 | 50 | Filtered |