Window level presets for CT data: ReferenceTopic number: 1501609936566

Two types of window level presets are available for CT data: linear and filtered. A linear preset is defined by center and width values and a filtered preset is defined by a combination of an available linear preset and an image filter.

Window level presets are defined by the following criteria:

PresetDescription
NameThe name of the defined preset.
Window CenterA numeric value defining the center of the window. This value can be positive or negative.
Window WidthA numeric value defining window width. This value must be positive.

The following presets are defined by default for CT data:

NameWindow WidthWindow CenterType
Lung1500-500Linear
Mediastinum15088Linear
Soft tissue35050Linear
Liver15085Linear
Bone2000500Linear
Abdomen/Pelvis30070Linear
Brain8020Linear
Post Fossa12547Linear
Bone + sharpen filter2000500Filtered
Lung + sharpen filter1500-500Filtered
Soft tissue + smooth filter35050Filtered